»ó¼¼¼³¸í

Performance
Measurement TechnologyThermo-paramagnetic oxygen sensor
GasProcess and non-condensing sample
Measurement RangeSelectable from 0-1% to 0-25% oxygen and 90- 100% (other ranges available on request)
Accuracy<¡¾1% of span for all ranges except ¡¾2% for 0-1% and 0.2% O2 for 90-100%
Response Time (T90)Standard response: 70 seconds
EN 50104: 45 seconds
Repeatability¡¾0.2% of span
Linearity¡¾0.5% of span for ranges greater than 0 - 5% 
<¡¾1% of span for ranges less than or equal to 0 - 5%
Zero Stability¡¾0.25% of span per month
Span Stability¡¾0.25% of span per month
Sample Flow Rate0.3 l/min for the sensor cell
Maximum Sample PressureAtmospheric to maximum of 3 barg (43 psig)
Maximum Sample Temperature55¡ÆC max (45¡ÆC for standard cell)
Electrical Output/Input
Output SignalOne isolated (galvanic ) 4–20 mA output (for O2 concentration)
Output Ranges0-1%; 2%; 5%; 10%; 25%; 90-100%
AlarmsTwo concentration alarms (open collector)
Display Resolution0.01%
Power Supply24 V DC; 1.5 A max
Operating Conditions
Ambient Temperature Range0¡ÆC to +55¡ÆC. (45¡ÆC for standard cell)
Mechanical Specification
Warm-up Time1 hour
Dimensions180 x 140 x 213mm (8.4 x 5.6 x 7.1 in) (h x w x d)
Allow 100mm on all sides for gas connections and cable entries
Weight6.3kg (13.9 lbs)
Wetted MaterialAISI 316 and Viton¢ç. Other materials available on request
Gas Connection¨ù" NPT female
Enclosure Ingress ProtectionIP66  Flame arrestors are supplied for hazardous area applications
GENERAL
Hazardous Area Certifications II 2 GD Exd IIC T4 Gb 
 II 2 D Extb IIIC T135¡ÆC Db IP66 T Amb -50¡ÆC/+60¡ÆC